The Zenith Branch Endovascular Graft-Iliac Bifurcation with the H&L-B One-Shot™ Introduction System is indicated for the endovascular treatment of patients with an aortoiliac or iliac aneurysm, an insufficient distal sealing site within the common iliac artery, and having morphology suitable for endovascular repair, including:
Adequate iliac/femoral access compatible with a 20 French (7.7 mm O.D.) introduction system.
Non-aneurysmal external iliac artery fixation segment distal to the aneurysm; with a length of at least 20 mm and with a diameter measured outer wall to outer wall of no greater than 11 mm and no less than 8 mm.
Non-aneurysmal internal iliac artery segment distal to the aneurysm: with a length of at least 10 mm (with 20-30 mm being preferred) and with a diameter acceptable for proper sealing.
- Zenith® Branch with Spiral-Z™: Brochure View document
- Physician's Pocket Reference Guide View document
- Zenith® Branch Deployment Sequence Presentation View document
- Zenith® Branch with Spiral-Z® Guidelines for Planning and Sizing View document
- Zenith Branch with Alpha Spiral-Z Guidelines for Planning and Sizing View document
- Zenith Branch with Alpha Spiral-Z Iliac Leg Planning and Sizing Worksheet View document
- Zenith® Branch with Spiral-Z® Device Planning and Sizing Worksheet View document
- Instructions for use View document